A-scan Specifications: 10 MHz transducer (nominal), IOL formulas: SRK/T, Binkhort II, SRK II, Holladay I, Haigis, Hoffer Q.
B-scan Specifications: 12.5 MHz or 20 MHz transducer; focus, 21 mm ± 2mm.
| 12.5 MHz Transducer | 20 MHz Transducer | |
|
Depth of Field |
14mm-37mm | 15mm-35mm |
| Axial Resolution | <0.35µs |
<0.13µs |
UBM Specifications: 35 MHz or 50 MHz transducer, focal point: 13mm (nominal), depth of field: 11.5-14mm.
| 35 MHz Transducer | 50 MHz Transducer | |
| Axial Resolution | 65µm |
50µm |
Minimum Computer Specifications: Microsoft Windows 7, 8, 8.1, or 10 ; 1024 x 768 pixel display; 2 GB RAM ; 1 GB hard drive (100 GB recommended) ; USB port for Scanmate Flex;
USB port for footswitch; mouse or touchpad; medical grade power supply.
A-scan Specifications: 10 MHz transducer (nominal), IOL formulas: SRK/T, Binkhort II, SRK II, Holladay I, Haigis, Hoffer Q.
B-scan Specifications: 12.5 MHz or 20 MHz transducer; focus, 21 mm ± 2mm.
| 12.5 MHz Transducer | 20 MHz Transducer | |
|
Depth of Field |
14mm-37mm | 15mm-35mm |
| Axial Resolution | <0.35µs |
<0.13µs |
UBM Specifications: 35 MHz or 50 MHz transducer, focal point: 13mm (nominal), depth of field: 11.5-14mm.
| 35 MHz Transducer | 50 MHz Transducer | |
| Axial Resolution | 65µm |
50µm |
Minimum Computer Specifications: Microsoft Windows 7, 8, 8.1, or 10 ; 1024 x 768 pixel display; 2 GB RAM ; 1 GB hard drive (100 GB recommended) ; USB port for Scanmate Flex;
USB port for footswitch; mouse or touchpad; medical grade power supply.